Maintenance 142
5
8. Attach the assembled in-line filters as indicated in Figure 5-9.
9. Reconnect the power cord, and turn on the autosampler.
10. Flush the autosampler at 1 mL/min for 10 minutes.
11. Inspect the in-line filters for leaks, and tighten 1/4-turn as required.
12. Replace the autosampler cover (see Section 2.6.1,
Removing/Replacing the
717plus Cover).