Appendix A SCPI Command Reference
MEASure subsystem
WCDMA Options Version 6.20
325
:MEAS[:CONT]:AFAN:DISTortion
:MEASure[:CONT]:AFANalyser:SINad
:MEASure[:CONTinuous]:PSUPply:GROup
Syntax
:MEAS[:CONT]:AFAN:DISTortion
Parameters
There are no parameters.
Description
Starts a continuous measurement of the third-harmonic distortion of a sine wave
applied to the audio analyzer. To read out the latest measurement result, use the
:FETCh:AFAN:DIST command.
Note: Any AF test command needs the Audio Option to be installed on your Willtek
4400.
Query
The query form of this command starts the measurement and delivers a string, con-
taining a floating point real number, representing the first measurement result value.
The physical dimension is percent (%).
Example
:MEAS:CONT:AFAN:DISTortion?
String delivered back: "0.3".
Syntax
:MEASure[:CONT]:AFANalyser:SINad
Parameters
There are no parameters.
Description
Starts a continuous measurement of the SINAD (i.e. signal to noise ratio). To read out
the latest measurement result, use the :FETCh:AFAN:SIN command.
Note: Any AF test command needs the Audio Option to be installed on your Willtek
4400.
Query
The query form of this command starts the measurement and delivers a string, con-
taining one floating point real number, representing the first measurement result
value. The physical dimension is dB.
Example
:MEAS:CONT:AFANalyser:SINad?
Value delivered back: "50.5".
Syntax
:MEASure[:CONTinuous]:PSUPply:GROup
Parameters
There are no parameters.
Description
Starts a continuous measurement of the power supply tests specified with the
:CONF:MEAS:GRO:PSUP command. To read out the latest measurement
results, use the :FETCh:PSUP:GROup command.
Notes
To perform any power and current consumption measurements on your Willtek
4400, the MS Power Supply and Current Measurements Option must be installed.
For further details regarding group measurements, please refer to section
:MEASure:...:GROup.