Appendix A SCPI Command Reference
CALCulate Subsystem
WCDMA Options Version 6.20
379
:CALCulate:AFANalyser:MSIGma?
:CALC:AFANalyser:ALL:LIMit[:FAIL]?
Parameters
There are no parameters.
Description
There is solely a query form of this command available.
Query
Returns the maximum measurement result value of the Audio measurement started
last. The string delivered back will contain as many maximum values as measurement
types performed. The single maximum values will always have the format of floating
point real numbers and will be separated by commas.
Example
:CALC:RES
:MEAS:ARR:AFAN:ALL
:CALC:AFAN:MMAX?
Syntax
:CALCulate:AFANalyser:MSIGma?
Parameters
There are no parameters.
Description
There is solely a query form of this command available.
Query
Returns the mean value and the standard deviation of the audio measurement
results since the last measurement start. The string delivered back contains as many
pairs of mean and standard deviation values as measurement types performed. The
individual values all have the format of floating point real numbers and will be sepa-
rated by commas.
Example
:CALC:RES
:MEAS:AFAN:ACV
...(other SCPI commands)...
:CALC:AFAN:MSIG?
In this example, first a continuous measurement of the audio voltage is started. After
some time, the :CALC:AFAN:MSIGma command is used to read out the mean
voltage and its standard deviation. The string returned in this example is "1.1,
0.2".
Syntax
:CALC:AFANalyser:ALL:LIMit[:FAIL]?
Parameters
There are no parameters.
Description
There is solely a query form of this command available.