CELL-DYN 3000 System Operator’s Manual 5-15
9140240E — May 1995
Chapter 5 Operating Instructions
7.Press [RETURN] to return to the QC SETUP MENU Screen.
NOTE: The X-B status is shown on the RUN Screen or
DISPLAY SPECIMEN Screen below the Data Station Status
Box (for example, X-B: 10/IN or X-B: 13/OUT2). The number
of specimens in the current X-B batch is noted (1-19). The status
of the previous batch is displayed following this number. The
status may be OUT, OUT2,
or IN.
•OUT indicates that the mean of the previous batch was outside
the action limits.
•OUT2 indicates that the mean of the last 2 or more completed
batches fell outside the action limits.
•IN indicates that the mean of the previous batch was within the
action limits.
The feature enabling the transfer of QC limits and data into a QC file
from a floppy disk is not currently available. Therefore the following
keys mentioned in this manual should not be used at this time: [LAB ID
SETUP], [LOAD FROM DISK], and [WRITE QC TO DISK].
LAB ID
SETUP