6-20 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Calibration Chapter 6
The [START AUTO-CAL] key starts the Auto-Cal program by
initiating three background counts.
The [CONTINUE AUTO-CAL] key is displayed after the background
counts are completed as shown in Figure 6.7.
NOTE: The key label changes to [INTERRUPT AUTO-CAL]
after the key is pressed.
The WHOLE BLOOD AUTO-CAL RESULTS Screen (see Figure6.8)
and the following soft key labels are displayed when the [CONTINUE
AUTO-CAL] key is pressed:
INTERRUPT AUTO-CAL
QUIT AUTO-CAL
PRINT
RETURN
The [INTERRUPT AUTO-CAL] key is used to interrupt (pause) the
Auto-Cal program.
The [QUIT AUTO-CAL] key is used to exit from the Auto-Cal program
before it is completed. When the [QUIT AUTO-CAL] key is pressed,
the bulletin line displays the message: ALL EXISTING DATA AND
RESULTS WILL BE CLEARED. The following soft key labels are
displayed:
CONFIRM QUIT
CANCEL QUIT
These keys are used to confirm or cancel the Quit Auto-Cal command.
The [PRINT] key is used to print the WHOLE BLOOD AUTO-CAL
RESULTS Screen.
The [RETURN] key is used to return to the reference value entry screen.
START
AUTO-CAL
CONTINUE
AUTO-CAL
INTERRUPT
AUTO-CAL
QUIT
AUTO-CAL
PRINT
RETURN