CELL-DYN 3000 System Operator’s Manual 6-19
9140240E — May 1995
Chapter 6 Calibration
Figure 6.6: Whole Blood Auto-Cal Screen
The [CLEAR REF VALS] key is used to clear the reference values used
in previous calibrations before entering new values. The following soft
key labels are displayed when the [CLEAR REF VALS] key is pressed:
CONFIRM CLEAR
CANCEL CLEAR
These keys are used to confirm or cancel the Clear Reference Values
command.
The [PRINT SUMMARY] key is used to print the entered reference
values.
The [RETURN] key is used to return to the AUTO-CALIBRATION
Screen.
EDIT
REF VAL
START
AUTO-CAL
QUIT
AUTO-CAL
CLEAR
REF VALS
PRINT
SUMMARY
RETURN
WHOLE BLOOD AUTO-CAL
Ready
May 11 1992
Operator ID
Sequence #
10:01
734
4100
Open Sampler
Enter reference value for each parameter to be calibrated: (up to 10 specimens)
Spec ID # of runs WBC RBC HGB MCV PLT
(1 to 10) (2.0-25.0) (2.00-6.50) (4.0-24.0) (70.0-100.) (50.0-600.)
CAL#01 1 7.30 4.24 12.9 90.0 244.
CAL#02 3 ---- ---- ---- ---- ----
CLEAR
REF VALS
PRINT
SUMMARY
RETURN