CELL-DYN 3000 System Operator’s Manual 6-21
9140240E — May 1995
Chapter 6 Calibration
Figure 6.7: Whole Blood Auto-Cal Screen
Figure 6.8: Whole Blood Auto-Cal Results Screen
EDIT
REF VAL
CONTINUE
AUTO-CAL
QUIT
AUTO-CAL
CLEAR
REF VALS
PRINT
SUMMARY
RETURN
WHOLE BLOOD AUTO-CAL
Ready
May 11 1992
Operator ID
Sequence #
10:01
734
4100
Open Sampler
Enter reference value for each parameter to be calibrated: (up to 10 specimens)
Spec ID # of runs WBC RBC HGB MCV PLT
(1 to 10) (2.0-25.0) (2.00-6.50) (4.0-24.0) (70.0-100.) (50.-600.)
CAL#01 3 7.30 4.24 12.9 90.0 244.
CAL#02 3 ---- ---- ---- ---- ----
INTERRUPT
AUTO-CAL
QUIT
AUTO-CAL
PRINT RETURN
WHOLE BLOOD AUTO-CAL
Ready for Calibration
May 11 1992
Operator ID
Sequence #
10:30
734
4100
Open Sampler
RESULTS FOR SPECIMEN #1
Spec 1 Mean Factor
Parameter Value RUN1 RUN2 RUN3 Factor Factor %Diff
WBC 7.30 ---- ---- ---- ---- ---- ----
RBC 4.24 ---- ---- ---- ---- ---- ----
HGB 12.9 ---- ---- ---- ---- ---- ----
MCV 90.0 ---- ---- ---- ---- ---- ----
PLT 244. ---- ---- ---- ---- ---- ----
SPEC ID CAL#01-01
NO. OF RUNS 3