CELL-DYN 3000 System Operator’s Manual 7-11
9140240E — May 1995
Chapter 7 Quality Control
When the key is pressed, the key label changes to [ACCEPT
SPECIMEN], an “R” is displayed in the column immediately left of the
results and the statistics are recomputed excluding those results. (See
Figure7.6.) The data is still displayed and stored in the file but is
excluded from the statistical calculation.
When the [ACCEPT SPECIMEN] key is pressed, the “R” is deleted
and the statistics are recomputed including those results.
Figure 7.6: View QC Log Screen (With Rejected Results)
The [DELETE SPECIMEN] key is used to delete the results for the
specimen indicated by the cursor position. When the [DELETE
SPECIMEN] key is pressed, the following key labels are displayed:
CONFIRM DELETION
CANCEL DELETION
NOTE: These keys are used to confirm or cancel the delete
command. When the [CONFIRM DELETION] key is pressed,
the results are deleted from the file (the data is not displayed or
stored in the file) and the statistics are recomputed excluding
those results.
Upper Limits: 8.3 4.58 13.8 89.5 17.6 265. 10.3
Lower Limits: 7.1 4.22 13.0 83.5 13.2 215. 8.3
Seq WBC RBC HGB MCV RDW PLT MPV Date Time Op
1838 7.6 4.37 13.1 85.9 15.6 221. 8.8 O 03/02/93 09:50 734
1839 7.5 4.31 13.0 85.9 15.8 224. 9.3 O 03/02/93 09:52 734
1840 7.7 4.33 13.1 85.8 15.1 232. 8.8 O 03/02/93 09:52 734
1937 7.3 4.38 13.1 85.9 15.5 239. 9.8 O 03/03/93 09:06 734
1938 7.4 4.38 13.3 85.9 15.8 258. 8.8 O 03/03/93 09:07 734
1939 7.5 4.38 13.2 85.9 15.4 233. 9.0 O 03/03/93 09:08 734
2036 7.4 4.39 13.1 86.1 15.2 225. 8.8 O 03/04/93 08:36 734
2037 7.1 4.39 13.0 86.0 15.5 238. 8.6 O 03/04/93 08:37 734
2038 R 7.2 4.33 13.0 86.1 15.4 235. 9.1 O 03/04/93 08:38 734
2184 7.4 4.39 13.1 86.0 15.9 234. 8.9 O 03/05/93 09:15 734
WBC RBC HGB MCV RDW PLT MPV
N: 89 89 89 89 89 89 89
Mean: 7.5 4.35 13.2 85.8 15.6 234. 9.0
Std. Dev: 0.2 0.05 0.1 0.4 0.3 9. 0.2
CV (
.
/.): 2.3 1.3 1.1 0.5 1.6 3.9 2.6
PURGE
QC LOG
LEVEY-
JENNINGS
ACCEPT
SPECIMEN
DELETE
SPECIMEN
WRITE QC
TO DISK
PRINT
QC LOG
RETURN
Mar 15 1993
Operator ID
Sequence #
09:59
734
2723
Norm 45 OPEN: 90/120
Page 7 of 9
VIEW QC LOG
Ready
USE < OR > FOR MORE DATA