Service Manual Installation
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
123
7.13
Verification of exposure parameters
The exposure parameters (kV, time and dose) can be checked using two different
methods:
invasive method
based on the measurement of the test points on HF board
(require the use of multimeter and oscilloscope for time) This method is typical
used during verification done by technical service engineers
non-invasive method
based on measurement with Dose meter. This is the
typical method used by Phisician to verify periodically the unit
In order to make easier the exposure parameters measurements, X-MIND prime
3D has a dedicated modality that allows X-ray exposure without rotating the arm
and without exposure parameters modulation that typically occurs in a standard
exam.