Service Manual Troubleshooting
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
173
9.2.5
Error with code from E500 to E505
This range of errors are dedicated to MCU PC ethernet communication problems
due to incompatibility between ASP software and MCU firmware version and/or
ethernet hardware issues).
1.
check the ethernet connection and the network card settings (see paragraph
7.7.1).
2.
Power ON the unit and wait the connection to the PC-GUI. Verify the
compatibility between MCU firmware and ASP versions: update/downgrade the
FW-SW to a released/compatible configuration.
Note
Contact Technical Service to verify that the firmware and software versions
are compatible with the unit configuration.
Technical Service additional information required
:
-
Software logs
-
MCU SD card log (see paragraph 11.2.1).