Figure 47
1.
Verify the presence of the correct calibration files (8 files named as the SN of
the sensor mounted on the unit) in the calibration folder
C:\ProgramData\Acteon
Imaging\Panoramic X-Mind Prime Ceph\Calibration
and that all the calibration
options in the image processing menu of the in PhD_C_Test.exe are checked
(see paragraph 10.2).
2.
Perform tests of Error E760 (paragraph 9.2.7.11).
3.
Verify the integrity of tubehead X57 connector, pin 5 and 6.