SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
19
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
10 ms 0 % a 0°,
45°, 90°, 135°,
180°, 225°, 270°,
315°
20 ms 0% a
500 ms 70% a
5 s 0%
IEC 60601-1-2
Test level
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
X-MIND prime 3D requires
continued operation during
power mains interruptions,
it is recommended that X-
MIND prime 3D be powered
from an uninterruptible
power supply or a battery
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
IEC 60601-1-2
Test level
Power frequency magnetic
fields should be at levels
characteristics of a typical
location in a typical
commercial or hospital
environment
2.4
Cybersecurity measures
Like all computer-based systems, X-MIND prime 3D might be exposed to
Cybersecurity threats.
X-MIND prime 3D is equipped with hardware provisions that make sure that no
unwanted X-ray exposure, laser radiation or motorized movements can be
activated even in case of cyber-attack or software failure.
Nevertheless, in order to minimize the possibility of cyber-attacks, it is user
responsibility to make sure that the following protection measures are followed.
The initial software installation and system set-up shall be done by authorized
and trained personnel only and using the software provided with the machine.
Any software or firmware upgrade of the equipment shall be done by
authorized and trained personnel only.
After any software or firmware upgrade, or any other maintenance operation,
image quality checks shall be performed to ensure the system is working as
expected. Instructions are given in User Manual, chapter 7.
Password-protect each user account on the Windows login. Passwords shall be
strong enough (at least made of 8 alphanumeric characters), shall be safely
managed by every user (for example they have not been written down), and
should be periodically changed (if the system is supplied with a PC, the
Windows user is password-protected, but it is user responsibility to change the
default password and set new ones for all the different users that will have
access to the system).
Activate a screensaver that requires a password to be unblocked after a
timeout of 5-10 minute, giving this way an automatic timed method to
terminate sessions, preventing an unauthorized access to the computer when
it is not used (if the system is supplied with a PC, the screen saver is activated
by default).