Service Manual Installation
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
132
7.14.1
Table of pre-set anatomic parameters
Default exposure values in 2D Panoramic modes
Adult Patient
(14 seconds)
Child Patient
(12.8 seconds)
kV
mA
kV
mA
Small
76
9
66
8
Medium
80
9
68
8
Large
82
9
70
8
Default exposure values in 2D Sinus mode
Adult Patient
(9 seconds)
Child Patient
(9 seconds)
kV
mA
kV
mA
Small
68
8
64
8
Medium
72
8
66
8
Large
74
8
68
8
Default exposure values in 2D TMJ mode
Adult Patient
(10,6 seconds)
Child Patient
(10,6 seconds)
kV
mA
kV
mA
Small
70
8
64
8
Medium
74
8
66
8
Large
78
8
68
8
Exposure values in 3D Full Dentition, 3D Extended Volumes and 3D Airways
Adult Patient
(7 seconds)
Child Patient
(7 seconds)
kV
mA
kV
mA
Small
84
4
64
6.3
Medium
84
5
66
6.3
Large
84
6
68
6.3