Service Manual Troubleshooting
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
158
9.2
Displayed messages
The X-MIND prime 3D operative states and any detected errors are signalled by the
different activation of the three keyboard LEDs (see User Manual keyboard
description) and by the displayed operational and error messages on the PC
interface-G.U.I. (Graphical User Interface):
Operational messages
: are instructions which guides the operator in the
correct use of the unit.
Error messages
: are displayed by the GUI and describe the last occurred error.
There are two kind of errors messages:
1.
Messages that require a reset by clicking on OK button on the GUI and by
pressing the >0< button on the unit keyboard.
2.
Messages that can only be reset after the turning OFF and ON of the unit.
The error messages are divided into different areas that can be distinguished by
the error number; the following table contains the different errors with meanings.
Main MCU board
Code
Error description
Reference
paragraph
001 / 003
Internal MCU errors
9.2.1
500 ÷ 505
MCU Ethernet errors
9.2.5
MCU EEPROM configuration
Code
Error description
Reference
paragraph
100 / 101
Configuration area parameter doesnt match the
expected one
9.2.2.1
102
Wrong version number in configuration area
9.2.2.2
103 / 104
Timeout error occurred during an Eeprom
erase/write operation
9.2.2.3
Rotation motor
Code
Error description
Reference
paragraph
200
Zero position optical sensor of rotation axis
always activated
9.2.3.1
201
Zero position optical sensor never activated
9.2.3.1