9.2.4
Errors with code from E300 to E399
9.2.4.1
E360: RX button pressed on start-up or before exam
This message is displayed if, during the power ON phase or before starting of the
exam, one of the connected X-ray button, has been sensed as pressed.
1.
Verify if one of the X-ray buttons was intentionally / unintentionally pressed:
switch OFF the unit and release the button. Switch ON the unit and verify if the
issue is solved.
2.
Switch OFF and ON the unit, press the X-ray button and verify that the LED H11
on MCU board (A1) light-up according to the X-ray button activation: if is not
ok, verify the connected X-ray buttons and their connections.
If they are not ok, replace or fix the buttons and verify if the issue is solved. If
the error is still present, replace the MCU board (see paragraph 11.3.2).
9.2.4.2
E362: X-ray button released during emission
The above error message is displayed if the X-ray button is unintentionally /
intentionally released during an exam; the emission is stopped and all motors
released in order to allow the patient’s exit.
Verify if the X-ray button has been intentionally / unintentionally released during
the exam:
a.
If it was intentionally released, press button >0< to reset the error on the unit
and close the error window displayed on the GUI.
b.
If it was unintentionally released, refer to Error E360.