Service Manual - Description
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
24
3.
CLEANING AND DISINFECTION
In order to guarantee a good level of hygiene and cleaning, it is necessary to carry
out the following procedures.
Warning
Disconnect the unit from the mains
before performing any cleaning.
Do not let water or other liquids penetrate the unit, as these could cause corrosion
or short circuits.
Use only a wet cloth and a mild detergent to
clean the painted surfaces, accessories and
connection cables and then wipe with a dry
cloth; do not use corrosive, abrasive solvents
(alcohol, benzine, trichloroethylene).
The centring bite or the bite protective sleeve and the cephalometric ear pin sleeves
must be replaced after each exam.
Thoroughly clean the chin support, resting handles, temple clamps, ceph rods,
nasion reference and carpus plate whenever they are used.
The chin support, resting handles temple clamps, ceph rods, nasion reference and
carpus plate should be disinfected (when considered necessary) with a solution of
2% glutaraldehyde.
Note
To ensure a greater level of hygiene the handles of the equipment are
covered with a special antibacterial paint which, thanks to the emission of
silver ions, prevents the development of micro-organisms.