SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
15
2.3
Information about Electromagnetic Compatibility
Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
the accompanying documents.
Portable and mobile RF communications equipment can affect medical electrical
equipment.
The equipment can be installed both in professional buildings (e.g. hospitals or
clinics) and in residential buildings. Residential buildings, according to IEC 60601-1-
2 4th edition, are intended to be connected to dedicated power supply system
(normally fed by separation transformers).
For the purpose of EMC environment classification according to IEC 60601-1-2 4th
edition, both installations are classified as "Professional healthcare facility
environment".
The EMISSIONS characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment, even
if it is usually permanently installed in X-Ray shield locations, might not offer
adequate protection to radio-frequency communication services. If abnormal
performance is observed, such as degradation of essential performance in the form
of lack of accuracy of exposure parameters and lack of reproducibility of exposure
parameter, additional measures may be necessary, such as re-orienting or
relocating the device.
Warning
The use of cables other than:
Ethernet cable CAT 6 L=5 m - code 5007090100
Ethernet cable CAT 6 L=10 m - code 5007090300
with the exception those sold by the manufacturer of the equipment or
system as replacement parts for internal components, may result in
increased emission or decreased immunity of the equipment or system.
Warning
X-MIND prime 3D should not be used adjacent to or stacked with other
equipment; if adjacent use is necessary, X-MIND prime 3D has to be
observed to verify if it operates in a normal way.
Interference may occur in the vicinity of equipment marked with the
symbol .