Service Manual – Troubleshooting
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
9.2.7
Error with code from 750 to 770
Warning
Those errors are related to the X-ray generator, so they can be safety
related.
In case of Error messages E759 and E755, the system must be immediately
powered off, because an unexpected emission (E759) can be present or the
emission has not been terminated into the expected time.
Warning
On the Generator board (A2) there are dangerous high voltage, 230 VAC /
120 VAC and 400 VDC.
Before accessing the Generator board, it is mandatory to switch OFF the
unit, disconnect it from the mains and wait up to 4 minutes in order to allow
the discharge of the capacitor (LED H10 on the Generator board steady
OFF).