Service Manual Corrective maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
234
11.2
Checks, settings and adjustment
11.2.1
Logs files recover
The X-MIND prime 3D firmware and software record some of the events that occurs
during the unit functioning, stored in files called Logs. These files have to be
provided to the Technical Service as required by the different error descriptions.
The following table lists all the logs file names and their path location, while the
paragraph below explains the procedures for activating and collecting them.
File name
Path location
Active by
default
Software
package
LogsServer_yyyy.mm.dd.l
og
C:\ProgramData\Acteon
Imaging\LogServer\Logs
Yes
MCU
eeprom.dump
mcu.log
[SDCARD]:\mcu\Logs
NO
CCU
CCU.log
[SDCARD]:\ccu\Logs
NO
3D
Sensor
dcam_yyyy.mm.dd-
hh.mm.ss.log
GigE_yyyy.mm.dd-
hh.mm.ss.log
C:\ProgramData\Acteon
Imaging\LogServer\Logs in the sub-
folders
“Logs_dcam“ and “Logs_GigE“
NO
Ceph
Sensor
C:\Program Files (x86)\Teledyne
DALSA\Sapera\Bin\logview.exe
Yes