Service Manual Corrective maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
230
11.1.2
CCU Firmware upgrade
1.
Copy in a working directory of the PC the CCU firmware to upload (the file name
is in the form CCU-YYYY-MM-DD-vM.mm.bbb.hex).
2.
Power ON the machine. Open "PhD_C_Test.exe" service program (C:\Program
Files (x86)\Acteon Imaging\Panoramic X-Mind Prime Ceph).
3.
Click on SETUP button and in the Login windows that will open insert the
password
PhdAccess
and click OK
4.
From the menu
Settings
select
Firmware update
>
Automatic mode
>
CCU
.
5.
In the Login window that will open insert the password
EthUpload
and click
OK.