Service Manual Description
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
17
2.3.1
Electromagnetic emissions
In accordance with the IEC 60601-1-2 Ed4 standard, X-MIND prime 3D is suitable
for use in the electromagnetic environment specified below.
The customer or user of the system must ensure that it is used in the said
environment.
Emissions test
Electromagnetic environment
RF emissions
CISPR 11
X-MIND prime 3D uses RF energy only for
its internal function. Therefore, its R.F.
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
X-MIND prime 3D is suitable for use in all
establishments other than domestic and
those directly connected to the public low
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonics emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3