Service Manual Corrective maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
312
11.3.3
A11 CCU board replacement
Replacement of this board doesn’t require any adjustment. Take care of cable
connections.
11.3.4
A2 Generator board replacement
Replacement of this board doesn’t require any adjustment. Take care of cable
connections Refer to paragraph 11.3.9 steps 1-4 for more information.
11.3.5
A12 Collimator driver board replacement
Replacement of this board doesn’t require any adjustment. Take care of cable
connections.
11.3.6
A13 CEPH driver board replacement
Replacement of this board doesn’t require any adjustment. Take care of cable
connections.
11.3.7
A14 Ceph sensor board replacement
Replacement of this board doesn’t require any adjustment. Take care of cable
connections.
11.3.8
Ceph sensor replacement
Once replaced the CEPH sensor, make a complete centering check as described in
paragraph 7.12.
11.3.9
3D sensor replacement
Note
Before a 3D sensor replacement, it is MANDATORY to:
Properly set the network interface board (see paragraph 7.7.2).
Activate the sensor logs (see paragraph 11.2.1.4) and try to reproduce
the error
Provide the sensor logs* (see paragraph 11.2.1.4) to Technical Service

Table of Contents