Service Manual - Description
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
18
2.3.2
Electromagnetic immunity
In accordance with the IEC 60601-1-2 Ed4 standard, X-MIND prime 3D is suitable
for use in the electromagnetic environment specified below.
The customer or user of the system must ensure that it is used in the said
environment.
Immunity
test
IEC 60601-1-2
test level
Compliance
level
Electromagnetic
environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
2/4/8/15 kV air
IEC 60601-1-2
Test level
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Radiated
electromagnetic
field
IEC 61000-4-3
3 V/m
80 MHz to 2.7
GHz
IEC 60601-1-2
Test level
Portable and mobile RF
communications equipment
should be used no closer to
any part of X-MIND prime
3D including cables.
Minimum distance 30 cm
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
> 3 m
IEC 60601-1-2
Test level
Mains power quality should
be that of a typical
commercial or hospital
environment
Surge
IEC 61000-4-5
0.5/1 kV
differential mode
0.5/1/2 kV
common mode
IEC 60601-1-2
Test level
Mains power quality should
be that of a typical
commercial or hospital
environment
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
150 kHz to 80
MHz
6 V
ISM frequencies
IEC 60601-1-2
Test level
Portable and mobile RF
communications equipment
should be used no closer to
any part of X-MIND prime
3D, including cables.
Minimum distance 30 cm