SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
11
2.1
Warnings
The device must be used in compliance with the procedures described and never
be used for purposes different from those herewith indicated.
Before performing any maintenance operation, disconnect the unit from the power
supply using the provided circuit breaker.
X-MIND prime 3D is an electro-medical device and therefore it can be used only
under the supervision of suitably qualified medical personnel, with the necessary
knowledge on X-ray protection.
The user is responsible for the fulfilment of the legal requirements regulating the
ownership, installation and use of the equipment itself.
This device has not been designed to be used in environments where vapours,
anaesthetic mixtures flammable with air, or oxygen and nitrous oxide, can be
detected.
Do not let water, or other liquids, into the device, as this could cause short-circuits
and corrosion.
Before cleaning the device, be sure that the main power supply has been
disconnected from the equipment. Pushing the ON/OFF button of the equipment,
it mustn't switch on.
Wherever necessary, use the appropriate accessories, such as the leaded aprons,
to protect the patient from radiation.
While performing the radiography, no-one, apart from the operator and the
patient, must remain in the room.
X-MIND prime 3D has been built to support a continuous operation at intermittent
load; therefore please follow the described use cycles to enable the device to cool
down.
X-MIND prime 3D must be switched off while using electrosurgical devices or
similar apparatus.
Warning
For safety reasons, it is prohibited to abnormally overload the patient
support arm, for example by leaning on it. The traction force on the handle
shall be less than 16kg.
Warning
To avoid risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
Please clean and disinfect, when necessary, all parts that can be in contact with the
patient.
The centering bite or the bite protective sleeve and the ear pin covers must be
replaced after each examination in which they were used.
Never try to rotate the moving arm manually when the unit is switched on, to avoid
permanent damage to the unit.
Movement is only possible in case of Error 362 because motors are disabled to
permit the patient exit.