Service Manual – Service programs descriptions
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
In order to enter in Configuration menu, type the password “TechAccess” in “Access
to Setup menu” filed and press Enter.
This page is reserved only to authorized technicians: it allows access to the different
functional parameters, as following:
•
Network Setting:
allows to set the IP address of the unit (see paragraph 8.1)
•
Exposition:
allows to disable the x-ray emission permanently (see paragraph
8.2)
•
Logs:
this page displays the exam counters (see paragraph 8.3)
Each time a parameter is modified, the unit will provide a confirmation window.