- Ceph sensor
In case one specific axis does not move, refer to the specific error related to
that axis:
- Primary collimator->E650-E661
- Secondary collimator->E640-E645
- Ceph sensor->E630-635
9.2.6.6
E606 Flat Panel is open, but should be closed
This error is shown when the Flat Panel sensor is not closed in 3D/pan position while
9.2.6.7
E623: timeout error during the CCU EEPROM erase operation /
E624: timeout error during the CCU EEPROM write operation
These errors are shown for a timeout during the CCU EEPROM erase or write
operation.
Power OFF the unit and, after 1 minute delay, power it ON again and verify the
correct functioning of the unit, checking the machine configuration and the set-up
data.
In case the error is persistent report it to the technical service.
If a new error is displayed, refer to the specific error paragraph description to fix
the issue).
.