EasyManua.ls Logo

Acteon X-MIND prime 3D ceph - Page 272

Acteon X-MIND prime 3D ceph
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual Corrective maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
270
7.
Open the image in SyMage software application (C:\Program Files
(x86)\Acteon Imaging\Panoramic X-Mind Prime Ceph\SyMage)
8.
In the menu bar select the “Rectangle Selection” tool
9.
Verify that the that the misalignment between the centres of the inner dot
and the outer circle is not more than 15 pixel (0.099mm/pixel) both in vertical
and horizontal directions:
|H1-H2|≤15pixels
|W1-W2|≤15pixels
Adjustments
Vertical ear rods corrections:
Warning
The vertical regulation of the ear rods will affect also the projection of the
secondary collimator on image plane, which will vary the upper and lower
not exposed borders position. Therefore after this adjustment repeat the
tests images requested by paragraph 7.12.

Table of Contents

Related product manuals