SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
26
between 60kV and 86kV, in 2kV steps and anodic current from 2 mA to 12.5 mA
in the R20 scale steps.
3D TMJ (FOV 85 x 93 mm) with two different FOV positions (R or L), 3 sizes for a
total of 12 combinations with automatic selection; with manual selection, it is
possible to select a high voltage between 60kV and 86kV, in 2kV steps and
anodic current from 2 mA to 12.5 mA in the R20 scale steps.
3D Sinus (FOV 85 x 93 mm) with 3 sizes for a total of 6 combinations with
automatic selection; with manual selection, it is possible to select a high voltage
between 60kV and 86kV, in 2kV steps and anodic current from 2 mA to 12.5 mA
in the R20 scale steps.
3D Extended Volumes (FOV 116 x 103 mm) with 3 sizes for a total of 6
combinations with automatic selection; with manual selection, it is possible to
select an high voltage between 60 kV and 86 kV, in 2 kV steps and anodic
current from 2 mA to 12.5 mA in the R20 scale steps.
3D Airways (FOV 116 x 103 mm) with 3 sizes for a total of 6 combinations with
automatic selection; with manual selection, it is possible to select a high voltage
between 60 kV and 86 kV, in 2 kV steps and anodic current from 2 mA to 12.5
mA in the R20 scale steps.
Cephalometric L-L projections in the formats 18x24, 24x24, 30x24 and 18x18, 24x18,
30x18; the selection between HS High Speed and HD High Definition is available.
Cephalometric A-P projections in the format 24x24 and 24x18 the selection between
HS High Speed and HD High Definition is available.
Carpus Projection in the format 18x24, only in HD High Definition mode.
Note of cephalometric image formats:
For user convenience, the ceph projections are named following the conventional format of
the film-cassettes (24 cm), although the vertical active area of the cephalometric sensor is
22.8 cm.