6240B DC Voltage Current Source/Monitor Operation Manual
4.2.6 MENU Key (Parameter Setting)
4-14
Header Switches the header ON or OFF.
On: Header ON
Off: Header OFF
Talk Only Switches between Addressable and Talk Only.
NOT displayed when the USB interface is selected.
On: Talk Only
Off: Addressable
SYSTEM Sets the 6240B system parameters.
Limit Buz Buzzes when the limiter (compliance) is activated.
On: Turns ON the limiter detection buzzer.
Off: Turns OFF the limiter detection buzzer.
CompareBuz Buzzes according to the comparator calculation result of measure-
ment data.
Off: Turns OFF the comparator calculation buzzer.
HI: Sounds when the comparator calculation result is HI.
GO: Sounds when the comparator calculation result is GO.
LO: Sounds when the comparator calculation result is LO.
HI or LO:
Sounds when the comparator calculation result is HI or
LO.
Notice Buz Buzzes when operations such as memory full, saving or loading
parameters are complete.
On: Turns ON the notice buzzer.
Off: Turns OFF the notice buzzer.
Self Test Executes the self-test for selected items.
Self Test: Executes the self-test.
Dsp/KeyTst: Executes the display and key test.
Error Log Can read the data stored in the error log when ERR indicator turns
ON.
Displays the number of errors that occurred, error codes and error
messages.
Once this parameter is displayed, the error log data will be cleared
and the ERR indicator goes OFF.
Relay Cnt Displays the operation count of the Operate/Standby relay.