6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-28
Sweep
2-slop linear
sweep
SM fd: First value
0.01 mV/
0.001 A
[fd, md,
ld, st1, st2]
md: Middle value
1 mV/
0.1 A
ld: Last value
2 mV/
0.2 A
st1: 1st step value
0.01 mV/
0.001 A
st2: 2nd step value
0.01 mV/
0.001 A
If all the settings are omitted, set the sweep type only.
However, it is not allowed to omit each value sepa-
rately.
SM? Response: SM<fd>, <md>,<ld>,<st1>,<st2>
*1
fd,md,ld,,st1,st2: <d.ddddEd>
Sweep type SX? The sweep type of the present source function is
returned.
Response: Linear sweep:
Same as the SN? response
Fixed level sweep:
Same as the SF? response
Random sweep:
Same as the SC? response
2-slope linear sweep:
Same as the SM? response
*1: The decimal point position in response varies depending on the setting value.
For the source value, limit value, and time parameter setting ranges, refer to the specifications.
Item Command Description
Default Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADCMT 6240B and is the answer not in the manual?

ADCMT 6240B Specifications

General IconGeneral
BrandADCMT
Model6240B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals