6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-30
Sweep
Sweep range SR0 Auto
SR1 Fixed
SR? Response: SR0 or SR1
Reverse mode SV0 OFF
SV1 ON
SV? Response: SV0 or SV1
Sweep repeat
count
SS cnt cnt: Count (0 to 1000)
(0 indicates infinite loop.)
1
SS? Response: SSdddd
Sweep stop SWSP Stop sweep
Trigger *TRG Sweep trigger
Measurement trigger
Mea
sure
ment
Function F0 Measurement OFF
F1 DC voltage measurement (DCV)
F2 DC current measurement (DCI)
F3 Resistance measurement (OHM)
F? Response: F0 to F3
Measurement
range
R0 AUTO range
R1 Fixed to the limit value range
(However, if the measurement function and the source
function are same, the measurement range conforms to
the source range.)
R? Response: R0 or R1
Measurement
function link
mode
FX0 OFF
FX1 ON (VSIM/ISVM)
(However, if measurement is set to OFF, the measure-
ment OFF status remains even after this parameter is
changed.)
FX? Response: FX0 or FX1
Item Command Description
Default Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS