6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-33
Cal
cula
tion
NULL calculation NL0 OFF
NL1 ON
NL? Response: NL0 or NL1
KNL  data Sets the NULL constant. (An error occurs when the
NULL calculation is set to OFF.)
*4
0
KNL? Response: KNL  d.dddddE  dd
Comparator
calculation
CO0 OFF
CO1 ON
CO? Response: CO0 or CO1
KHI  data Sets the upper limit.
0
KLO  data Sets the lower limit.
*4
0
KHI? Response: KHI  d.dddddE  dd
KLO? KLO  d.dddddE  dd
Scaling calcula-
tion
SCL0 OFF
SCL1 ON
SCL? Response: SCL0 or SCL1
KA a a: Constant A (0 (zero) is invalid.)
1
KB b b: Constant B
0
KC c c: Constant C
*4
1
KA? Response: KA  d.dddddE  dd
KB? KB  d.dddddE  dd
KC? KC  d.dddddE  dd
MAX/MIN
calculation
MN0 OFF
MN1 ON
MN? Response: MN0 or MN1
AVE? Reads out the average value.
0
MAX? Reads out the maximum value.
-9.99999E+26
MIN? Reads out the minimum value.
+9.99999E+26
TOT? Reads out the total value.
0
AVN? Reads out the measurement count.
Response: AVN d.dddddE+dd
0
*4: The setting range is from 0 to  999.999E + 24.
Item Command Description
Default Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS