6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-34
System
User parameter STP0 Saves the current setting parameters to area 0 of the
non-volatile memory.
STP1 Saves the current setting parameters to area 1 of the
non-volatile memory.
STP2 Saves the current setting parameters to area 2 of the
non-volatile memory.
STP3 Saves the current setting parameters to area 3 of the
non-volatile memory.
SINI Sets the factory default settings in all the areas from 0
to 3.
RCLP0 Loads the data from area 0 of the non-volatile memory
as setting parameters.
RCLP1 Loads the data from area 1 of the non-volatile memory
as setting parameters.
RCLP2 Loads the data from area 2 of the non-volatile memory
as setting parameters.
RCLP3 Loads the data from area 3 of the non-volatile memory
as setting parameters.
RINI Loads the factory default settings as setting parameters.
Initialization *RST Initializes parameters.
(The parameters except marked by
*6
in this list are the
factory default values.)
C Device clear
Instrument
information
*IDN? Tells the instrument to ask itself.
Response: ADC Corp.nnnnn,xxxxxxxxx,yyyyy
ADC Corp.: manufacturer (9 characters)
nnnnn: model
"6240B" (5 characters)
xxxxxxxxx: serial number (9 characters)
yyyyy: ROM revision (5 characters)
Power frequency Auto setting
LF? Response: LF0...50 Hz
LF1...60 Hz
Notice buzzer NZ0 OFF
NZ1 ON
NZ? Response: NZ0 or NZ1
Item Command Description
Default Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS