6240B DC Voltage Current Source/Monitor Operation Manual
9.1 Source and Measurement
9-5
Typical value of full scale item tolerance
The full-scale item tolerances listed below are added to the integration time 100 s to 10 ms and the 1-day
stability.
S/H: Measurement in the sample hold mode (integration time: 100 s)
When LO OUTPUT is grounded to the chassis, the additional error of integration time in 30 A range is the
same as that in 300 A range.
Source linearity: 3 digits or less (5 digits or less in the 4 A range)
Maximum output current: 0 to 15 V; 1 A (DC)
0 to 10 V; 4 A (maximum pulse width 20 ms/duty factor 20 %
Maximum compliance voltage: Up to 1 A (DC); 0 to 15 V
Up to 4 A (pulse); 0 to 10 V
Voltage source range Voltage limiter range A Rm B Is Tolerance
30 mA 300 mV 10 V
10 m
3 A 10 mA 1 m
300 mA 300 mV 10 V 30 A 100 mA 103
1 A 300 mV 10 V 300 A 1 A 13
30 mA 300 mV 10 V
100 m
3 A 10 mA 1 m
300 mA 300 mV 10 V 30 A 100 mA 130
1 A 300 mV 10 V 300 A 1 A 40
30 mA 300 mV 10 V
1
3 A 10 mA 1.3 m
300 mA 300 mV 10 V 30 A 100 mA 400
1 A 3 V 50 V 300 A 1 A 350
Meas
urement
range
Integration time Unit: digits (at 5
1
/
2
digit display)
10 ms 5 ms 1 ms 500 s100 sS/H
Voltage
measurement
300 mV 10 15 20 30 60 200
3 V 5 8 10 15 30 50
15 V 5 8 10 15 30 50
Current
measurement
30 A 200 300 300 300 300 300
300 A 2030303070100
3 mA 10 30 30 30 50 80
30 mA 10 30 30 30 50 100
300 mA 10 15 15 15 50 100
1 A 20307575250500
4 A 20307575250500