6240B DC Voltage Current Source/Monitor Operation Manual
2.2.8 Random Sweep Source
2-48
Preparation
1. Initialize the 6240B following the same procedure described in Section 2.2.5,
"DC Measurement."
Setting the sweep source mode
2. Press MODE. Rotate   or  press   or   to select DC-SWP, and press
ENTER.
The DC indicator and the SWP indicator go ON, showing the DC sweep has been
set.
3. Press MENU.
4. Press  or  to select B) SWEEP. Press   to enter the Select level.
5. Rotate   or press   or   to select Random.
Setting the sweep source range.
6. Press   and   to select 2) SWP Range. Press   to enter the Input/Run
level.
7. Press  to select Fix. Single-range sweep is set.
Setting the start and stop addresses
8. Press  ,   and   in order to select C) SWEEP VAL.
9. Press  to select 1) Sweep Adr. Press   to enter the Input/Run level.
10. Press NULL (SEL) to select the stop address.
The start address is displayed at half-brightness, and the last digit of the stop
address blinks.