6240B DC Voltage Current Source/Monitor Operation Manual
2.2.8 Random Sweep Source
2-50
18. Press 123..., 2, and ENTER in order.
The period time is set to 2 ms.
Setting measurement to OFF
19. Press , and in order to select E) MEASURE.
20. Press , and in order to select 3) Measure SW. Press to enter
the Input/Run level.
21. Press to select Off.
Setting the random memory
22. Press , , and in this order to select G) RANDOM MEM.
23. Press to select 1) Data Set... Press to enter the Input/Run level.
24. Press 123..., 1, 0, 0, 0 and ENTER in order.
1 V is set at address 0, and address 1 is displayed.
25. 25. Press 123..., 2, 0, 0, 0 and ENTER in order.
2 V is set at address 1, and address 2 is displayed.
26. Press 123..., 3, 0, 0, 0 and ENTER in order.
3 V is set at address 2, and address 3 is displayed.
To check the entered values, press NULL (SEL) and rotate counterclock-
wise.
27. Press MENU.
The HOME screen is displayed.