2-88 Chapter 2
System Verification and Performance Tests
Agilent 8753ET System Verification and Performance Tests
Initialization
1. Clear all internal memory.
CAUTION This will erase all instrument states that may be stored in internal memory.
Perform the following steps to save any instrument states that are stored in
internal memory to a floppy disk.
a. Press .
b. Select an instrument state and press .
c. Press .
d. If the instrument state file was not saved to disk with the same name that it had
while in internal memory, you may wish to rename the file.
Press , enter the desired name, and press
.
e. Repeat steps a through d for each instrument state that you wish to save.
To clear all internal memory, press
.
2. Connect the equipment as shown in Figure 2-49. Let the system warm up for 30
minutes.
Figure 2-49 System Verification Test Setup
Save/Recall
System
Preset