Chapter 2 2-149
System Verification and Performance Tests
Agilent 8753ES Performance Test Records
Agilent Technologies Company
Model 8753ES
Serial Number
Report Number
Option(s) Date
7. Test Port Input Frequency Response
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Frequency Range Test Port Specification
(dB)
Measured Value
(dB)
Meas.
Uncert.
(dB)
300 kHz–3 GHz Port 2 ± 1 0.47
300 kHz–3 GHz Port 1 ± 1 0.47
3 GHz–6 GHz Port 1 ± 2 0.17
3 GHz–6 GHz Port 2 ± 2 0.17