2-148 Chapter 2
System Verification and Performance Tests
Agilent 8753ES Performance Test Records
Agilent Technologies Company
Model 8753ES
Serial Number
Report Number
Option(s) Date
6. Test Port Input Noise Floor Level
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Frequency
Range
Test Port IF
Bandwidth
Specification
(dBm)
a
a. Specifications in parentheses are valid for analyzers with Option 075.
Calculated
Value
Meas.
Uncert.
(dB)
300 kHz–3 GHz Port 1 3 kHz ≤ −82 (−80) ±2.0
300 kHz–3 GHz Port 1 10 Hz ≤ −102 (−100) ±2.0
300 kHz–3 GHz Port 2 10 Hz ≤ −102 (−100) ±2.0
300 kHz–3 GHz Port 2 3 kHz ≤ −82 (−80) ±2.0
3 GHz–6 GHz Port 2 3 kHz ≤ −77 ±2.0
3 GHz–6 GHz Port 2 10 Hz ≤ −97 ±2.0
3 GHz–6 GHz Port 1 10 Hz ≤ −97 ±2.0
3 GHz–6 GHz Port 1 3 kHz ≤ −77 ±2.0