2-170 Chapter 2
System Verification and Performance Tests
Agilent 8753ET Performance Test Records
Agilent Technologies
Company
Model 8753ET
Serial Number
Report Number
7. Transmission Test Port Input Noise Floor Level
Note: If your analyzer does not have Option 006, write “N/A” for entries above 3 GHz.
Frequency Range IF
Bandwidth
Calculated
Val ue
Specification
(dBm)
Measurement
Uncertainty
(dBm)
300 kHz to 3.0 GHz 3 kHz ≤ −90 ± 1.0
300 kHz to 3.0 GHz 10 Hz ≤ −110 ± 1.0
3 GHz to 6.0 GHz 10 Hz ≤ −105 ± 2.0
3 GHz to 6.0 GHz 3 kHz ≤ −85 ± 2.0