2-6 Chapter 2
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
Required Equipment
Procedure
Initialization
1. Clear all internal memory.
CAUTION This will erase all instrument states that may be stored in internal memory.
Perform the following steps to save any instrument states that are stored in
internal memory to a floppy disk.
a. Press .
b. Select an instrument state and press .
c. Press .
d. If the instrument state file was not saved to disk with the same name that it had
while in internal memory, you may wish to rename the file.
Press , enter the desired name, and press
.
e. Repeat steps a through d for each instrument state that you wish to save.
2. To clear all internal memory, press
.
3. Connect the equipment as shown in Figure 2-1. Let the system warm up for 30 minutes.
Description HP/Agilent Model Number
Calibration Kit: 7-mm 85031B
Verification Kit: 7-mm 85029B
Test Port Extension Cable Set: 7-mm 11857D
Printer HP ThinkJet/LaserJet
Additional Equipment for Option 075 Analyzers
Minimum-Loss Pad (2): 50 Ω to 75 Ω 11852B
Adapter (2): APC-7 to Type-N (m) 11525A
Save/Recall
System
Preset