EasyManuals Logo

Agilent Technologies 8753ET Service Guide

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #344 background imageLoading...
Page #344 background image
Chapter 7 7-7
Source Troubleshooting
Phase Lock Error
1. Make sure the A9 CC switch is in the ALTER position:
a. Remove the power line cord from the analyzer.
b. Set the analyzer on its side.
c. Remove the two corner bumpers from the bottom of the instrument with a T-15
TORX screwdriver.
d. Loosen the captive screw on the bottom cover’s back edge.
e. Slide the cover toward the rear of the instrument.
f. Set the switch to the ALTER position as shown in Figure 7-2.
g. Replace the bottom cover, corner bumpers, and power cord.
Figure 7-2 Switch Positions on the A9 CPU
2. Switch on the analyzer and press
to generate new analog bus correction constants. Then press
to generate default
pretune correction constants.
Press to
generate new pretune correction constants.
3. Press and observe the analyzer display:
No error message: set the A9 CC switch to the NORMAL position. Then refer to
“Post-Repair Procedures” on page 14-53 to verify operation.
Error message visible: continue with A4 Sampler/Mixer Check”.
System
SERVICE MENU
TESTS
46 x1
EXECUTE TEST
System
SERVICE MENU
TESTS
45 x1
EXECUTE TEST
System
SERVICE MENU
TESTS
48 x1
EXECUTE TEST
YES
Preset

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals