EasyManuals Logo

Agilent Technologies 8753ET Service Guide

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #363 background imageLoading...
Page #363 background image
7-26 Chapter 7
Source Troubleshooting
Phase Lock Error
Figure 7-24 High Quality Comb Tooth at 3 GHz
3. If the signal at the A7 output is good, check the A7-to-A4 cable.
4. If the signal is not as clean as Figure 7-24, observe the HI OUT input signal to the
A7 assembly.
a. On the network analyzer, press
. Otherwise do not readjust the instrument. Remove the A14-to-A7
SMB cable (W9) from the A7 pulse generator assembly (CW ≈ 16 MHz).
b. Set the spectrum analyzer to a center frequency of 45 MHz and a span of 30 MHz.
Connect it to the A14-to-A7 cable still attached to the A14 assembly. Narrow the
span and bandwidth to observe the signal closely.
5. If the HI OUT signal is as clean as Figure 7-25, the A7 assembly is faulty.
Otherwise, check the A14-to-A7 cable or recheck the A13/A14 fractional-N as described
ahead.
Rechecking the A13/A14 Fractional-N
Some phase lock problems may result from phase noise problems in the fractional-N loop.
To troubleshoot this unusual failure mode, do the following:
1. Set the network analyzer at 60 MHz in the FRACN TUNE mode.
2. Use a spectrum analyzer, to examine the HI OUT signal from the A14 assembly. The
signal should appear as clean as Figure 7-25. The comb shape may vary from pulse
generator to pulse generator.
System
SERVICE MODES
PLL AUTO OFF

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals