Chapter 2 2-15
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
Figure 2-8 Test Port Output Frequency Range and Accuracy Test Setup
2. Press .
3. Press and write the frequency counter reading on the performance test
record.
4. Repeat step 3 for each instrument frequency listed in the performance test record.
Preset
Sweep Setup
CW FREQ
30 k/m