EasyManuals Logo

Agilent Technologies 8753ET Service Guide

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
2-26 Chapter 2
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
4. Press .
5. Press
.
6. Press (or the next power setting from the performance test record).
7. Press .
8. Press
.
9. Read the value for marker 1 from the display and enter it in the performance test
record in the “Marker 1 Maximum” value column. Enter the value of marker 2 in the
“Marker 2 Minimum” value column.
10.Press
(or the next power setting from the performance test record) .
11.Repeat steps 6 through 10 for all power settings listed on the performance test record
for the 300 kHz to 3 GHz frequency range.
12.The marker 1 (maximum) and marker 2 (minimum) readings should be within the
specified range for each power level with reference to the fixed marker value.
13.If your analyzer does not have Option 006, this completes the test. If you have problems
with this test, go to “In Case of Difficulty” on page 2-27.
14.If your analyzer has Option 006, proceed to the next two sections.
Power Meter Calibration from 3 GHz to 6 GHz (Option 006 only)
1. Disconnect the 8482A power sensor from the power meter and connect the 8481A power
sensor in its place. Zero and calibrate the power meter for this sensor. Connect the
equipment as shown in Figure 2-11 on page 2-23, using the 8481A sensor.
2. Press .
3. Press .
4. Press . Press the appropriate keys to
build a sensor calibration factor table for the 8481A power sensor (sensor B).
5. Press to exit the sensor calibration factor entries menu.
6. Press to select the 8481A sensor. Press
.
7. Press to start the power
meter calibration.
8. Proceed to the next section for making the power linearity measurements.
Display
DATA MEMORY
DATA/MEM
Marker
MODE MENU
FIXED MKR POSITION
FIXED MKR STIMULUS
300 k/m
FIXED MKR VALUE
15
x1
RETURN
REF =
FIXED MKR
Power 15 x1
Sweep Setup
TRIGGER MENU
SINGLE
Marker
MARKER 1
Marker Search
SEARCH: MAX
Marker
MARKER 2
Marker Search
SEARCH:MIN
Marker
MODE MENU
FIXED MKR POSITION
FIXED MKR VALUE
13 x1
RETURN
Start 3 G/n Stop 6
G/n
Cal
PWRMTR CAL
0 x1
LOSS/SENSR LISTS
CAL FACTOR SENSOR B
DONE
USE SENSOR B
Sweep Setup
TRIGGER MENU
CONTINUOUS
Cal
PWRMTR CAL
ONE SWEEP
TAKE CAL SWEEP

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals