EasyManuals Logo

Agilent Technologies 8753ET Service Guide

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
2-64 Chapter 2
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
8. Set the step attenuator to 10 dB.
9. Set the following analyzer parameters:
•
•
•
•
10.Set up the 8753ES for power meter calibration:
a. Select the 8753ES as the system controller:
•
•
b. Set the power meter's address:
•
•
c. Select the appropriate power meter by pressing until the correct
model number is displayed. (Use the 438A/437 selection if the power meter is an
E4419B or E4418B.)
d. Select the calibration kit and enter the power sensor calibration data.
•
• (enter
the power sensor calibration data for 30 MHz)
11.Take a power meter calibration sweep.
•
•
12.Verify that the power meter reads approximately −20 dBm.
Full 2-Port Calibration
13.Connect the equipment as shown in Figure 2-35.
Preset
Sweep Setup
CW FREQ
30
M/µ
NUMBER of POINTS
51 x1
Power −10 x1
Avg
IF BW
10 x1
Local
SYSTEM CONTROLLER
SET ADDRESSES
ADDRESS: P MTR/GPIB
13
x1
POWER MTR
Cal
CAL KIT
SELECT CAL KIT
7mm 85031
Cal
PWRMTR CAL
LOSS/SENSOR LISTS
CAL FACTOR SENSOR A
DONE
Cal
PWRMTR CAL
−20 x1
ONE SWEEP
TAKE CAL SWEEP

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals