Page #408 background image
408 Chapter6
Error Messages
Error Message Descriptions
cdmaOne - Specific Error Messages
[200 to 299]
An error number in the range [200 to 299] indicates the instrument has
detected an error relating to the cdmaOne personality.
cdmaOne - Specific Error Message Descriptions
(Number) Description
(200) Signal near noise floor - Power accuracy
degraded
(201) Signal exceeds maximum allowable power -
Reduce input power
(202) Input overload
Excessive input power has been detected which will
cause the ADC to clip the signal. Reduce the signal
level, change the attenuator/max total power setting
(under
Input menu), or press Restart if the RF Input
Range is Auto.
(203) Channel center frequency outside device’s
transmit band
(205) No power at carrier frequency
No power was detected as a CW or a modulated signal.
(206) Cannot correlate to input signal
A correlation failure with the pilot CDMA channel
occurred during synchronous demodulation.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals