EasyManuals Logo

Agilent Technologies E4406A VSA Series User Manual

Agilent Technologies E4406A VSA Series
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #409 background imageLoading...
Page #409 background image
Chapter 6 409
Error Messages
Error Message Descriptions
NADC - Specific Error Messages
[300 to 399]
An error number in the range [300 to 399] indicates the instrument has
detected an error relating to the NADC personality.
NADC - Specific Error Message Descriptions
(Number) Description
(300) Sync word not found
In an EVM measurement, the sync word is not found
and the synchronization cannot be established when
Sync Word is selected in the Burst Sync menu.
(301) Valid NADC burst not found
A valid NADC burst is not found when the
Device is
MS.
(302) Signal too noisy
The valid EVM measurement cannot be performed,
because the input signal is too noisy.
(303) Burst Delay exceeds 2 ms limit for EVM
In an EVM measurement, the
Burst Delay value must be
less than 2 ms.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals