410 Chapter6
Error Messages
Error Message Descriptions
PDC - Specific Error Messages
[400 to 499]
An error number in the range [400 to 499] indicates the instrument has
detected an error relating to the PDC personality.
PDC - Specific Error Message Descriptions
(Number) Description
(400) Sync word not found
In an EVM measurement, the sync word is not found
and the synchronization cannot be established when
Sync Word is selected in the Burst Sync menu.
(401) Valid PDC burst not found
A valid PDC burst is not found when the
Device is MS.
(402) Signal too noisy
The valid EVM measurement cannot be performed,
because the input signal is too noisy.
(412) Burst Delay exceeds 2 ms limit for EVM
In an EVM measurement, the
Burst Delay value must be
less than 2 ms.