Chapter 4 161
Basic Digital Operation
Using Waveform Clipping
Configuring Rectangular Clipping
This procedure shows you how to configure rectangular clipping. The rectangular setting clips the I and Q
data independently. For more information about rectangular clipping, refer to “How Clipping Reduces
Peak-to-Average Power” on page 157.
1. Press
Preset > Mode > Custom > Arb Waveform Generator > Digital Modulation Off On to On. This generates a
custom arbitrary waveform for use in this procedure. You can also use a previously stored or
downloaded waveform.
2. Press
Mode > Dual ARB > Select Waveform and ensure that AUTOGEN_WAVEFORM is highlighted on the
display. AUTOGEN_WAVEFORM is the default name assigned to the waveform you generated in the
previous step.
3. Press
Select Waveform. This selects the waveform and returns you to the previous softkey menu.
4. Press
ARB Off On to On. The dual Arb player must be turned on to display the CCDF plot in the following
steps.
5. Press
ARB Setup > Waveform Utilities > Waveform Statistics and ensure that AUTOGEN_WAVEFORM is
highlighted on the display.
6. Press
CCDF Plot and observe the position of the waveform’s curve, which is the darkest line.
7. Press
Return > Return > Clipping.
8. Ensure that the
Clipping Type |I+jQ| |I|,|Q| softkey is set to |I|,|Q|. This activates the Clip |I| To and
Clip |Q| To softkeys that enable you to configure rectangular (independent) I and Q data clipping.
9. Press
Clip |I| To > 80 > %.
10. Press
Clip |Q| To > 40 > % > Apply to Waveform. The I and Q data are individually clipped by 80% and
40%, respectively. Once clipping is applied to the waveform it cannot be undone. Repeated use of the
clipping function has a cumulative effect on the waveform.
11. Press
Waveform Statistics > CCDF Plot and observe the waveform’s curve. Notice the reduction in
peak-to-average power, relative to the previous plot, after applying clipping.